Page 18 - The Beauty and Sorrow in Endodontics (WL and C/S)
P. 18

apical tissue, the circuit is completed, and the apex locator calibrates the measurement to

            0 at this location. As the file move back coronally, the machine will then calculate the
            distance from the tip of the file to the apex using a set of algorithm, which in reality, is

            still  just  an  estimate.  Furthermore,  some  operators  may  be  hesitant  to  peek  the  file

            through  the  apex  during  measurement  and  stops  when  the  apex  locator  signals  that
            0.5mm from the apex is reached; this result is highly inaccurate as the apex locator had

            not  been  calibrated  yet.  Therefore,  Dr.  Schilder  recommended  the  use  of  radiographic

            terminus  as  the  reference  point  for  working  length  for  the  following  benefits:  1.  It
            eliminates  the  potential  discrepancies  in  value  from  elongation  /  foreshortening  of  the

            films; and 2. No matter which angle the radiographs are taken at, the point at which the
            file goes beyond the radiographic terminus will always be the same.


            c.  Last WL; lWL:


                 Generally speaking, lWL is only used after cleaning and shaping for verification of

            the master GP cone during cone fit try-in. It is a variable number as the value can change
            until a desirable cone fit radiograph is obtained. The only rule is that the lWL will always

            be shorter than the iWL as canals will straighten up as they are shaped.

                 The term “lWL” is defined as the WL identified with the largest file in a series used

            to clean and shape the apical canal (even if said file is only used for size verification or as

            the master apical file). lWL is used to verify the length of the master cone prior to any
            cone fit radiographs. Once the master cone reaches the WL with adequate tugback, a plier

            is used to mark the outer reference point on the master cone. The cone is then removed

            from the canal and placed side by side with the master apical file for comparison (Fig
            A22), and if their lengths are reasonably similar, then the cone may now be placed back

            into the canal for a cone fit radiograph to be taken. If the lengths marked on the file and
            cone  are  obviously  different,  then  the  operator  needs  to  consider  the  following

            possibilities: either 1. the apical cross section diameter is different between the file and
            the cone; or 2. the tugback felt was not from the resistance in the apex, but from the cone

            binding to the wall in the coronal or middle thirds of the canal.


                 Since the horizontal section of the canal is never a perfect circle even after cleaning

                                                                                                               16
   13   14   15   16   17   18   19   20   21   22   23